Hall Probes Driver

Model: HPD 4.1

Plasma Diagnostics

 

 

 

Hall probe module 2301

Input amplifier gain

1×, 10×, 100×, 1000×

Total gain

10×, 100×, 1000×, 10000×

Input voltage

0 to ±0.5 V

Max output voltage

±0.5 V

High-pass filter

10 Hz

Low-pass filter

1 kHz, 10 kHz, 100 kHz, 350 kHz

Order of active low-pass filter

4

Output current

0 to 200 mA

Output current stability
(for ΔR = 50 W)

1%

Setting accuracy
(real value vs value on the potentiometer scale)

±1%

Standard accuracy
(main accuracy)

0.3%

Additional accuracy
(caused by temporal changes)

0.01 %/°C for total gain 10
0.01 %/°C for total gain 100
0.02 %/°C for total gain 1000
0.03 %/°C for total gain 10000

Setting time
(recommended time after power-on before exact measurement)

10 min

Temperature module 2302

Temperature probe

PT100

Temperature range

-200 to +200 °C

Calibrated range

0 to 100 °C

Output voltage

10 mV/°C and 0 V for 0 °C

Standard accuracy
(for 0 to 100 °C)

±0.3% ±1 digit

Additional accuracy

0.02 %/K

Communication module 2303

Please, ask the manufacturer for details.

Power supply module 2205

Power

230 VAC, 50/60 Hz, 20 VA

Dimmensions (system unit)

19” × 3U × 280 mm

Weight (system unit)

8 kg

 

Overview

|

Parameters

|

Schematics

|

Photos

|